PED. ARTERIAL CANNULA
Received Apr 17, 2026 · Event occurred Apr 1, 2026
Report 2184009-2026-00474 · MDR key 24907286
Device
Generic name
Catheter, Cannula And Tubing, Vascular, Cardio
Manufacturer
Medtronic Perfusion SystemsModel number
77008Catalog number
77008
Lot number
2024020594
Product problems
- Material Deformation
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ADDITIONAL INFO B5: MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE INFO THAT THE GUIDEWIRE CAUSED THE DEVICE TO FRAY IS CORRECT. INSERTION OF THE INTEGRATED GUIDE AND PLACEMENT OF THE NELATON CATHETER AT THE TIP. AT THAT STAGE, THE CANNULA BODY BEGAN TO FRAY OR UNRAVEL. THE FIBER DETACHMENT WAS SPECIFICALLY LOCATED IN THE SECTION BETWEEN POINTS 1 AND 2 IN THE DIAGRAM. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Additional Manufacturer Narrative
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULA, IT WAS REPORTED THAT UPON INSERTION OF THE GUIDEWIRE, THE CANNULA BODY DEFECTED, IMMEDIATELY FRAYING. IT WAS CONFIRMED THAT THE MATERIAL DID NOT COME INTO CONTACT WITH THE PATIENT ,AND THE DEFECT WAS DETECTED BEFOREHAND DURING TECHNICAL HANDLING.THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION B5. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE, THE DEVICE IS NOT SUBJECTED TO ANY TEMPERATURE CHANGES. E. INITIAL REPORTER. FIRST NAME, LAST NAME <(>&<)> PHONE NUMBER: THIS INFORMATION WAS ADDED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.