inforMED
InjuryEMN

HU-FRIEDY MFG CO., LLC.

Received Apr 16, 2026 · Event occurred Mar 20, 2026

Report 1416605-2026-00005 · MDR key 24903876

Device

Generic name

H6/h7 De Scaler, Ee2 #9 Handle

Model number

SH6/79E2

Catalog number

SH6/79E2

Lot number

0424

Product problems

  • Break

Patient

NA · Unknown

  • Foreign Body In Patient

Narrative

Additional Manufacturer Narrative

BREAKAGE DUE TO CRACK IN INSTRUMENT. THE OBSERVED CUTTING-EDGE DULLING AND LACK OF TIMELY RESHARPENING MAY HAVE CONTRIBUTED TO INCREASED FORCE APPLIED DURING CLINICAL USE, FURTHER PROMOTING CRACK PROPAGATION AND EVENTUAL FRACTURE. NO EVIDENCE OF MANUFACTURING OR MATERIAL DEFECT WAS IDENTIFIED.

Description of Event or Problem

CUSTOMER REPORTED THAT INSTRUMENT BROKE AND THE PATIENT STATED THEY SWALLOWED THE TIP/PIECE. PATIENT RECEIVED AN X-RAY AND THE INSTRUMENT PIECE WAS NOT VISIBLE ON X-RAY.