InjuryEMN
HU-FRIEDY MFG CO., LLC.
Received Apr 16, 2026 · Event occurred Mar 20, 2026
Report 1416605-2026-00005 · MDR key 24903876
Device
Generic name
H6/h7 De Scaler, Ee2 #9 Handle
Manufacturer
Hu-friedy Mfg. Co., LlcModel number
SH6/79E2Catalog number
SH6/79E2
Lot number
0424
Product problems
- Break
Patient
NA · Unknown
- Foreign Body In Patient
Narrative
Additional Manufacturer Narrative
BREAKAGE DUE TO CRACK IN INSTRUMENT. THE OBSERVED CUTTING-EDGE DULLING AND LACK OF TIMELY RESHARPENING MAY HAVE CONTRIBUTED TO INCREASED FORCE APPLIED DURING CLINICAL USE, FURTHER PROMOTING CRACK PROPAGATION AND EVENTUAL FRACTURE. NO EVIDENCE OF MANUFACTURING OR MATERIAL DEFECT WAS IDENTIFIED.
Description of Event or Problem
CUSTOMER REPORTED THAT INSTRUMENT BROKE AND THE PATIENT STATED THEY SWALLOWED THE TIP/PIECE. PATIENT RECEIVED AN X-RAY AND THE INSTRUMENT PIECE WAS NOT VISIBLE ON X-RAY.