EVIS LUCERA ELITE XENON LIGHT SOURCE
Received Apr 15, 2026 · Event occurred Mar 18, 2026
Report 3002808148-2026-11581 · MDR key 24889368
Device
Product problems
- Poor Quality Image
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: D8, G3, G6, H2, H3, H4, H6 AND H11. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE REPORTED FAILURE WAS NOT CONFIRMED. BASED ON THE RESULTS OF THE INVESTIGATION A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Description of Event or Problem
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Description of Event or Problem
IT WAS REPORTED THAT THE XENON LIGHT SOURCE EXHIBITED A NOISY IMAGE DURING INSPECTION FOR USE. THERE WAS NO PATIENT INVOLVEMENT.
Additional Manufacturer Narrative
THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.