inforMED
MalfunctionMVR

GEM MICROVASCULAR ANASTOMOTIC COUPLER

Received Apr 14, 2026

Report 1416980-2026-01289 · MDR key 24875092

Device

Generic name

Device,anastomotic,microvascular

Model number

GEM2754 I

Catalog number

511100300010

Lot number

SP25J09-2354065

Product problems

  • Difficult to Open or Close

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT FOUR (4) 3.0MM COUPLERS RING MAY OPEN OR FAIL TO MAINTAIN ITS PROPER CLOSURE DURING HANDLING, PREVENTING CORRECT COUPLING OF THE SYSTEM. THIS OCCURRED DURING SURGICAL PROCEDURES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Additional Manufacturer Narrative

B3/D10: THE EVENTS OCCURRED BETWEEN JANUARY TO FEBRUARY 2026. E1: INITIAL REPORTER PHONE NO.: (B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative

H11: A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. THE DEVICES WERE NOT RECEIVED FOR EVALUATION; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.