inforMED
MalfunctionHTJ

DEPTH GAUGE F/SCR Ø4.5 - 6.5 MEAS-RANGE

Received Apr 13, 2026 · Event occurred Jan 1, 2026

Report 8030965-2026-03381 · MDR key 24857000

Device

Generic name

Gauge, Depth

Manufacturer

Synthes Gmbh

Catalog number

319.100

Lot number

93250P1

Product problems

  • Incorrect Measurement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THE MEASURING RANGE CANNOT BE DISPLAYED BELOW 20MM. IT GETS STUCK. EVENT OCCURRED PRE-OPERATIVELY.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4) THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. INVESTIGATION SUMMARY THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THE DEVICE TO BE SLIGHTLY BENT FROM THE DISTAL END OF THE SLEEVE. A FUNCTIONAL TEST WAS PERFORMED WITH THE INTERACTION OF THE SLEEVE AND THE SCALE BODY AND PRESENTS SOME FRICTION WHEN YOU SLIDE THE SCALE BODY, THEREFORE THE REPORTED ALLEGATION CAN BE CONFIRMED. WITH AVAILABLE INFORMATION THE COMPLETE POTENTIAL CAUSE CANNOT BE ESTABLISHED. PROPERLY HANDLING AND ATTENTION TO THE APPROVED USE OF THE DEVICE DIMINISHES THE RISK OF FAILURE. A DIMENSIONAL INSPECTION WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE DEPTH GAUGE F/SCR Ø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