CLEAR-PB MOBILE BARRIERS
Received Mar 6, 2012 · Event occurred Feb 27, 2012
Report 2485616 · MDR key 2485616
Device
Generic name
Shield, X-ray, Mobile
Manufacturer
Fluke Biomedical (formerly Nuclear Associates)Model number
56-604Catalog number
*
Lot number
*
Product problems
- Break
- Component Falling
- Crack
Patient
67 YR
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
THE PATIENT WAS UNDERGOING CORONARY ANGIOGRAPHY. THE ATTENDING PHYSICIAN WAS WALKING ACROSS THE PROCEDURE ROOM IN ORDER TO INSERT SOME DOCUMENTS INTO THE PATIENT'S CHART. SINCE FLUOROSCOPY WAS ONGOING, HE WAS MOVING THE MOBILE X-RAY SHIELD ACROSS THE ROOM SO THAT HE WOULD REMAIN BEHIND IT. JUST AFTER HE LET GO, THE UPPER PORTION OF THE WINDOW SNAPPED OFF JUST ABOVE THE VERTICAL MOUNTS ON BOTH SIDES. THE CRACKED EDGE WAS SMOOTH BUT NOT SHARP LIKE GLASS. THE PIECE THAT FELL TO THE FLOOR DID NOT SHATTER NOR STRIKE ANYONE OR ANYTHING THAT WOULD HAVE COMPROMISED THE PROCEDURE.STAFF FORTUNATELY SAVED THE TWO PARTS OF THE BROKEN WINDOW. IT'S MADE OF PLASTIC, NOT GLASS SO THE PART THAT FELL DIDN'T SHATTER. THE DEVICE IS CONSIDERED A RADIATION SAFETY DEVICE BY THE MANUFACTURER. MY OPINION IS THAT THERE WAS A NOTCH NEAR ONE OF THE SIDE SUPPORT CHANNELS AND THAT THE CONTINUED FLEXING AND HANDLING OF THE UNSUPPORTED UPPER PORTION OF THE WINDOW OVER A LONG PERIOD OF TIME RESULTED IN A CRACK GROWING FROM THE NOTCH UNTIL THIS CATASTROPHIC EVENT OCCURRED.