inforMED
MalfunctionDRG

ORG-9110A

Received Apr 11, 2026 · Event occurred Apr 1, 2026

Report 8030229-2026-06451 · MDR key 24848867

Device

Generic name

Multiple Patient Receiver

Model number

ORG-9110A

Catalog number

ORG-9110A

Lot number

NA

Product problems

  • Communication or Transmission Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE MULTIPLE PATIENT RECEIVER (ORG) ORG WAS DROPPING OFF THE NETWORK RANDOMLY. HE SAID IT WILL NOT SHOW ON HOST TABLE AT ALL EVEN THOUGH ORG DOES NOT HAVE ERROR LIGHTS. THEY CONFIRMED IT IS NOT A NETWORK CABLE OR SWITCH ISSUE ALREADY BY TESTING DIFFERENT PORTS AND CABLES. THIS WOULD AFFECT THE ZM TRANSMITTERS. NO PATIENT HARM WAS REPORTED. NIHON KOHDEN CONTINUES TO INVESTIGATE THE REPORTED EVENT. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE MULTIPLE PATIENT RECEIVER (ORG) ORG WAS DROPPING OFF THE NETWORK RANDOMLY. HE SAID IT WILL NOT SHOW ON HOST TABLE AT ALL EVEN THOUGH ORG DOES NOT HAVE ERROR LIGHTS. THEY CONFIRMED IT IS NOT A NETWORK CABLE OR SWITCH ISSUE ALREADY BY TESTING DIFFERENT PORTS AND CABLES. THIS WOULD AFFECT THE ZM TRANSMITTERS. NO PATIENT HARM WAS REPORTED.