inforMED
MalfunctionJQX

BD PHOENIX¿ AP

Received Apr 9, 2026 · Event occurred Mar 10, 2026

Report 1119779-2026-00641 · MDR key 24835756

Device

Generic name

Nephelometer, For Clinical Use

Catalog number

448010

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED WHILE USING THE BD PHOENIX¿ AP THE USER STATED THAT THE INSTRUMENT WAS CONTAMINATED, CAUSING DISCREPANCIES IN ANTIBIOTIC RESISTANCE AND SENSITIVITY RESULTS ON THE DOWNSTREAM INSTRUMENT. THE AP CONSUMABLES HAVE BEEN CHANGED AND DOWNSTREAM RESULTS WERE IMPROVED. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Additional Manufacturer Narrative

H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.