inforMED
MalfunctionFLE

AMSCO

Received Apr 9, 2026 · Event occurred Mar 11, 2026

Report 3005899764-2026-00028 · MDR key 24831363

Device

Generic name

Steam Sterilizer

Catalog number

SR0201110311

Product problems

  • Pressure Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT AND FOUND THE VIEWPORT TO BE DAMAGED. THE TECHNICIAN REPLACED THE VIEWPORT, TESTED THE FUNCTION AND OPERATION OF THE AMSCO 400 STERILIZER, CONFIRMED IT TO BE OPERATING TO SPECIFICATIONS AND RETURNED IT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem

THE USER FACILITY REPORTED THAT DURING A CYCLE, THEIR AMSCO 400 STERILIZER MADE A "LOUD NOISE" AND STATED STEAM AND WATER WERE LEAKING FROM THE UNIT. NO REPORT OF INJURY.