FISHER & PAYKEL HEALTHCARE
Received Apr 9, 2026
Report 9611451-2026-01286 · MDR key 24825116
Device
Generic name
Infant Warmer
Manufacturer
Fisher & Paykel Healthcare LtdModel number
IW930AEUCatalog number
IW930AEU
Lot number
021004
Product problems
- No Audible Alarm
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
A HEALTHCARE FACILITY IN OKLAHOMA REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE IW930AEU SERVO-CONTROL COSYCOT INFANT WARMER FAILED THE POWER FAILURE ALARM TEST DURING A DEVICE MAINTENANCE. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Additional Manufacturer Narrative
(B)(6). SECTION D4: UDI DETAILS ARE NOT AVAILABLE BASED ON THE TIME OF MANUFACTURING. METHOD: THE SUBJECT DEVICE, IW930AEU SERVO-CONTROL COSYCOT INFANT WARMER WAS NOT RETURNED TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. OUR INVESTIGATION IS THUS BASED ON THE INFORMATION PROVIDED BY THE HEALTHCARE FACILITY, AND OUR KNOWLEDGE OF THE PRODUCT. RESULTS: THE HEALTHCARE FACILITY REPORTED THAT THE SUBJECT DEVICE FAILED THE POWER FAILURE ALARM TEST DURING A DEVICE MAINTENANCE CONCLUSION: WE ARE UNABLE TO DETERMINE THE CAUSE OF THE REPORTED EVENT. THE DEVICE TECHNICAL/SERVICE MANUAL CONTAINS A CHECKLIST WHICH SPECIFIES THAT USERS PERFORM SAFETY, PERFORMANCE, AND FUNCTIONAL CHECKS - INCLUDING THE POWER FAIL ALARM AT LEAST ONCE A YEAR. INCLUDED IN THE WARMER'S MAINTENANCE CHECKLIST IS A TEST OF THE SUPER CAPACITOR TO ENSURE IT IS OPERATING WITHIN SPECIFICATIONS.