inforMED
MalfunctionNUJ

NA

Received Apr 8, 2026 · Event occurred Jan 29, 2026

Report 0001056128-2026-00024 · MDR key 24822495

Device

Generic name

Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic,

Model number

LF4418

Catalog number

LF4418

Lot number

15787713

Product problems

  • Output Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THE DEVICE WAS "NOT WORKING." IT DIDN'T FIRE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO STRYKER SUSTAINABILITY SOLUTIONS FOR EVALUATION. THE RETURNED COMPLAINT DEVICE WAS RECEIVED WITH THE STRYKER SUSTAINABILITY PACKAGING AND NO TRAY. THERE WAS EVIDENCE OF BIO-BURDEN PRESENT ON THE DEVICE. THE HANDLE, BLADE TRIGGER, BUTTON, PLUG, AND JAWS APPEARED TO BE INTACT. THE DEVICE HAS A STRYKER PLUG, THE PLUG WAS INSPECTED FOR DAMAGE, CORROSION, AND DEFORMATION AND NONE WERE FOUND. THE SPACER PADS WERE INSPECTED FOR DAMAGE AND FOUND NONE. THUMB BUTTON FEEDBACK WAS ACCEPTABLE AS IT WAS ABLE TO DEPRESS THE PRESSURE PAD ON THE MEMBRANE SWITCH AND EXHIBITED A TACTILE CLICK, AND THE PRESSURE PAD DISENGAGED WHEN RELEASED. THE HANDLE FUNCTIONALITY WAS TESTED AND CONFIRMED TO BE ACCEPTABLE AS THE LEVER WAS ABLE TO ACTUATE THE JAWS MULTIPLE TIMES. HANDLE LEVER WAS RETURNED TO THE START POSITION AND THE JAWS OPEN WHEN RELEASED. AN AUDIBLE CLICK WAS HEARD WHEN THE JAW LEVER ENGAGED THE CLICK TAB. WHILE THE JAWS WERE IN THE LOCKED POSITION, THEY WERE INSPECTED AND FOUND TO BE PROPERLY ALIGNED. THE BLADE TRIGGER WAS TESTED AND VERIFIED TO MAINTAIN PROPER MOVEMENT. MANUAL CONTINUITY TESTS WERE PERFORMED AND FOUND NO CONNECTION TO THE CONTINUITY TEST ON PLUG PRONG #1 TO PRONG #2 WHEN THE ACTIVATION BUTTON IS ACTUATED. THE DEVICE WAS THEN DISASSEMBLED. INSPECTION OF THE SOLDER SLEEVES REVEALED NO EXPOSED WIRES. INSPECTION ISOLATED THE OPEN CIRCUIT AT THE ACTIVATION BUTTON ON THE SWITCH BOARD. INSPECTION OF THE BUTTON COVER, BUTTON AND CIRCUIT BOARD REVEALED NO IRREGULARITIES. THE CONTINUITY WAS TESTED ON THE PCB BUTTON AND EACH WIRE AND SOLDER SLEEVE. EACH WIRE AND SOLDER POINT TO THE PCB BOARD MADE A CONNECTION. CONTINUITY WAS RE-CHECKED ON PIN 1 AND PIN 2. BY PRESSING THE BUTTON WITH EXCESSIVE FORCE THE BUTTON MADE A CONNECTION. THE DEVICE INSPECTION AND FUNCTIONAL TEST INDICATED THAT THE COMPLAINT WAS CONFIRMED FOR THE REPORTED FAILURE MODE. A REVIEW OF THE DHR FOR THE REPORTED LOT/SERIAL NUMBER SUPPORTS THAT THE DEVICE MET ALL INSPECTION AND TEST CRITERIA PRIOR TO RELEASE FROM STRYKER THE REPORTED EVENT COULD BE ATTRIBUTED TO: - ELECTRICAL CONNECTIONS DAMAGE - DAMAGE DUE TO SHIPPING, HANDLING CONDITIONS. THE INSTRUCTIONS FOR USE (IFU) STATE: - USE CAUTION DURING SURGICAL CASES IN WHICH PATIENTS EXHIBIT CERTAIN TYPES OF VASCULAR PATHOLOGY (ATHEROSCLEROSIS, ANEURYSMAL VESSELS, ETC.). FOR BEST RESULTS, APPLY THE SEAL TO UNAFFECTED VASCULATURE - DO NOT USE THIS INSTRUMENT ON VESSELS IN EXCESS OF 7 MM IN DIAMETER. - PLACE THE VESSEL OR TISSUE IN THE CENTER OF THE JAWS. TO AVOID INCOMPLETE SEALING, DO NOT GRASP TISSUE BEYOND THE ELECTRODE SURFACE; DO NOT PLACE TISSUE IN THE JAW HINGE. - CONTACT BETWEEN AN ACTIVE INSTRUMENT ELECTRODE AND ANY METAL OBJECT (HEMOSTATS, STAPLES, CLIPS, RETRACTORS, ETC.) MAY INCREASE CURRENT FLOW AND MAY RESULT IN UNINTENDED SURGICAL EFFECTS, SUCH AS AN EFFECT AT AN UNINTENDED SITE OR INSUFFICIENT ENERGY DEPOSITION. - ELIMINATE TENSION ON THE TISSUE WHILE SEALING AND CUTTING TO ENSURE PROPER FUNCTION. - DO NOT ATTEMPT TO SEAL OVER CLIPS OR STAPLES AS INCOMPLETE SEALS/DAMAGE TO THE CUTTING BLADE WILL OCCUR. CONTACT BETWEEN AN ACTIVE ELECTRODE AND ANY METAL OBJECTS MAY RESULT IN ALTERNATE SITE BURNS OR INCOMPLETE SEALS. - A CONTINUOUS TONE SOUNDS TO INDICATE THE ACTIVATION OF RF ENERGY. WHEN THE ACTIVATION CYCLE IS COMPLETE, A TWO-PULSED SEAL-CYCLE-COMPLETE TONE SOUNDS AND RF OUTPUT CEASES. - THE SURGEON MAY INSPECT THE SEAL BEFORE CUTTING THE VESSEL OR TISSUE. AFTER INSPECTING THE SEAL, THE SURGEON SHOULD CREATE A SECOND SEAL ADJACENT TO THE FIRST SEAL BEFORE CUTTING, AS DESCRIBED BELOW. - KEEP THE INSTRUMENT JAWS CLEAN. BUILD-UP OF ESCHAR MAY REDUCE SEALING AND/OR CUTTING EFFECTIVENESS. WIPE JAW SURFACES AND EDGES WITH A STERILE, WET GAUZE PAD AS NEEDED. - A TONE WITH MULTIPLE PULSES INDICATES THAT THE SEAL CYCLE WAS NOT COMPLETED. DO NOT CUT TISSUE UNTIL YOU HAVE VERIFIED THAT THERE IS AN ADEQUATE SEAL. - USE CAUTION DURING SURGICAL CASES IN WHICH PATIENTS EXHIBIT INCOMPRESSIBLE TISSUE. THERE MAY BE LIMITATIONS ASSOCIATED WITH EFFECTIVE USE OF THE DEVICE IN THESE SITUATIONS. - ALTHOUGH THE JAWS OF THE INSTRUMENT WILL ACCOMMODATE TISSUE GREATER THAN 15 MM THICK, EXCEEDING THE TISSUE THICKNESS LIMIT OF 15 MM COULD RESULT IN COMPROMISED SEALS. - WHEN AN ALERT CONDITION OCCURS, ENERGY DELIVERY STOPS. AFTER THE ALERT CONDITION HAS BEEN CORRECTED, ENERGY DELIVERY WILL BE IMMEDIATELY AVAILABLE. ALERT SITUATIONS - WHEN AN ALERT CONDITION OCCURS, ENERGY DELIVERY STOPS, THE GENERATOR PRODUCES A SEQUENCE OF PULSED TONES, AND AN ALERT WILL BE DISPLAYED ON THE GENERATOR. DO NOT CUT THE VESSEL. THE USER SHOULD INSPECT THE SEAL SITE AND INSTRUMENT BEFORE PROCEEDING. AFTER THE ALERT CONDITION HAS BEEN CORRECTED, ENERGY DELIVERY WILL BE IMMEDIATELY AVAILABLE. TROUBLESHOOTING STEPS 1. RELEASE THE FOOTSWITCH PEDAL OR ACTIVATION BUTTON, IF STILL ENGAGED. 2. OPEN THE INSTRUMENT JAWS AND INSPECT FOR SUCCESSFUL SEAL. 3. FOLLOW THE SUGGESTED CORRECTIVE ACTIONS ON THE GENERATOR SCREEN, THE GENERATOR QUICK REFERENCE CARD, OR IN THE GENERATOR USER¿S GUIDE. 4. IF POSSIBLE, REPOSITION THE INSTRUMENT AND REGRASP TISSUE IN A LOCATION THAT OVERLAPS THE PREVIOUS SEAL, THEN REACTIVATE THE SEAL CYCLE. REASON FOR ALERT - TOO LITTLE TISSUE BETWEEN THE JAWS ¿ THE USER IS GRASPING THIN TISSUE OR NOT ENOUGH TISSUE; OPEN THE JAWS AND CONFIRM THAT SUFFICIENT AMOUNT OF TISSUE IS INSIDE THE JAWS. IF NECESSARY, INCREASE THE THICKNESS OF TISSUE THAT IS GRASPED AND REACTIVATE THE SEAL CYCLE. - TOO MUCH TISSUE BETWEEN THE JAWS ¿ THE USER IS GRASPING TOO MUCH TISSUE; OPEN THE JAWS, REDUCE THE AMOUNT OF TISSUE THAT IS GRASPED, AND REACTIVATE THE SEAL CYCLE. - ACTIVATING ON A METAL OBJECT ¿ AVOID GRASPING OBJECTS, SUCH AS STAPLES, CLIPS, OR ENCAPSULATED SUTURES IN THE JAWS OF THE INSTRUMENT. - DIRT JAWS ¿ USE A WET GAUZE PAD TO CLEAN SURFACES AND EDGES OF INSTRUMENT JAWS. - EXCESS FLUIDS IN SURGICAL FIELD ¿ MINIMIZE OR REMOVE EXCESS FLUIDS FROM AROUND THE INSTRUMENT JAWS. - ACTIVATION SWITCH RELEASED BEFORE SEAL COMPLETES TONE ¿ THE FOOTSWITCH OR ACTIVATION BUTTON WAS RELEASED BEFORE THE SEAL CYCLE WAS COMPLETE. - MAXIMUM SEAL CYCLE TIME HAS BEEN REACHED ¿ THE SYSTEM NEEDS MORE TIME AND ENERGY TO COMPLETE THE SEAL CYCLE. THE REPORTED EVENT WILL CONTINUE TO BE MONITORED THROUGH POST-MARKET SURVEILLANCE.