HF-RESECTION ELECTRODE "PLASMALOOP ¿ MEDIUM, 30°", 24 FR., STANDARD, 12°-30°
Received Apr 8, 2026 · Event occurred Mar 9, 2026
Report 3007960282-2026-00063 · MDR key 24821143
Device
Generic name
Hf-resection Electrode
Manufacturer
Olympus Medical Products Czech, Spol. S R.o.Model number
WA47506SLot number
NI
Product problems
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT DURING A HYSTEROSCOPIC SUBMUCOSAL MYOMECTOMY, THE ELECTRODE BROKE. IT WAS REPORTED THAT NO PART OF THIS DEVICE FELL INTO THE UTERUS. THE PROCEDURE WAS COMPLETED WITH ANOTHER ELECTRODE DEVICE, RESULTING IN A DELAY OF ABOUT 30 MINUTES. REPORT 2 OF 2.
Additional Manufacturer Narrative
E1: ESTABLISHMENT NAME: (B)(6) HOSPITAL. E1: ESTABLISHMENT ADDRESS: (B)(6). THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND CONFIRMED THE REPORTED PROBLEM: A VISUAL INSPECTION FOUND THAT THE LOOP ELECTRODE AT THE TIP HAD BROKEN. THE CAUSE OF THE LOOP ELECTRODE BREAKAGE CANNOT BE IDENTIFIED, BUT INSPECTION OF THE FRACTURE SURFACE SUGGESTS THAT IT MAY HAVE BEEN CAUSED BY OVERLOAD. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: COMPONENT FAILURE. ADDITIONAL INFORMATION WAS REQUESTED FROM THE CUSTOMER, BUT NO FURTHER INFORMATION HAS BEEN RECEIVED AT THIS TIME. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.