inforMED
MalfunctionFEM

COUPLER AC FFL 19

Received Apr 8, 2026 · Event occurred Mar 26, 2026

Report 1221934-2026-01449 · MDR key 24820370

Device

Generic name

Accessories, Photographic, For Endoscope (exclude Light Sources)

Catalog number

242437

Product problems

  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. H11: ADDITIONAL NARRATIVE: THE PRODUCT HAS NOT RETURNED TO DEPUY SYNTHES; HOWEVER A PHOTO WAS PROVIDED FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED PHOTO FOUND THE COUPLER DELAMINATED. NO OTHER ANOMALY COULD BE OBSERVED. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE COUPLER WOULD HAVE CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, A POTENTIAL CAUSE CANNOT BE ESTABLISHED. MULTIPLE FACTORS ARE ASSOCIATED WITH THIS TYPE OF FAILURE, AND THE COMPLAINT DEVICE NEEDS TO BE PHYSICALLY RECEIVED AND EVALUATED IN OUR SERVICE CENTER TO DETERMINE A POTENTIAL CAUSE OF WHY THE CUSTOMER EXPERIENCED THE FAILURE. IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.

Description of Event or Problem

IT WAS REPORTED THAT DURING THE PRE-OPERATION SCOPE PROCEDURE, IT WAS OBSERVED THAT THE COUPLER EXHIBITED METAL FLAKING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY REPLACING THE AFFECTED COUPLER. THERE WERE NO SURGICAL DELAYS OR HARM TO THE PATIENT AS A RESULT OF THIS INCIDENT. NO FURTHER DETAILS WERE AVAILABLE REGARDING THE INCIDENT.