MalfunctionKHP

I-STAT CG4+ CARTRIDGE

Received Apr 8, 2026 · Event occurred Mar 23, 2026

Report 2245578-2026-00087 · MDR key 24814828

Device

Generic name

Cg4+ Cartridge

Model number

NA

Catalog number

03P85-25

Lot number

M25317

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem

ON (B)(6) 2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CG4+ CARTRIDGE THAT YIELDED A SUSPECTED DISCREPANT RESULTS ON A PATIENT. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD: I-STAT DATE: (B)(6) 2026 TESTED: 19:04, PCO2: 51.2 MMHG, LACTATE: 2.26 MMOL/L SAMPLE: A. METHOD: I-STAT DATE: (B)(6) 2026 TESTED: 19:20 PCO2: 48.1 MMHG LACTATE: 9.88 MMOL/L SAMPLE: B. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.