Received Apr 8, 2026
Report 9611451-2026-01264 · MDR key 24812559
Device
Generic name
Infant Resuscitator
Manufacturer
Fisher & Paykel Healthcare LtdModel number
RD900AEUCatalog number
RD900AEU
Lot number
100923
Product problems
- Pressure Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
(B)(6). SECTION D4, UDI: UDI DETAILS ARE NOT AVAILABLE BASED ON THE TIME OF MANUFACTURING. METHOD: THE SUBJECT DEVICE, RD900AEU NEOPUFF INFANT RESUSCITATOR WAS RECEIVED AT OUR FISHER & PAYKEL HEALTHCARE (F&P) SERVICE CENTRE IN CALIFORNIA WHERE IT WAS PERFORMANCE TESTED BY A TRAINED F&P SERVICE TECHNICIAN. OUR INVESTIGATION IS THUS BASED ON THE INFORMATION PROVIDED BY THE F&P SERVICE TECHNICIAN, THE INFORMATION PROVIDED BY THE HEALTHCARE FACILITY AND OUR KNOWLEDGE OF THE PRODUCT. THE SERVICE CENTRE ALSO REPORTED THAT THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER. RESULT: UPON EVALUATION OF THE SERVICE REPORT, IT WAS OBSERVED THAT THE NEEDLE OF THE SUBJECT DEVICE'S MANOMETER DOESN'T FUNCTION PROPERLY AND READY INACCURATELY. CONCLUSION: BASED ON THE EVALUATION OF THE SERVICE REPORT AND OUR KNOWLEDGE OF THE PRODUCT, THE REPORTED EVENT COULD HAVE RESULTED FROM THE DEVICE BEING SUBJECTED TO AN IMPACT DAMAGE. THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF TECHNICAL MANUAL WARNS AGAINST DROPPING THE NEOPUFF OR SUBJECTING IT TO IMPACT DAMAGE WHICH MAY CAUSE THE UNIT TO OPERATE INCORRECTLY. IF THE NEOPUFF IS SUSPECTED TO HAVE BEEN DAMAGED, THE MANOMETER AND VALVE SYSTEM SHOULD BE PERFORMANCE TESTED. IN ADDITION, THE NEOPUFF USER INSTRUCTIONS STATE THAT THE USER SHOULD "CHECK MANOMETER READS ZERO WITH NO GAS FLOW" AND CHECK THE PRESSURE SETTINGS "PRIOR TO EVERY USE OF THE NEOPUFF". AS MENTIONED, THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: - DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT.
Description of Event or Problem
DURING DEVICE EVALUATION AND PERFORMANCE TESTING OF A RD900AEU NEOPUFF INFANT RESUSCITATOR AT OUR FISHER & PAYKEL HEALTHCARE (F&P) SERVICE CENTER IN CALIFORNIA, THE MANOMETER WAS FOUND TO BE FAULTY. THERE WAS NO REPORTED PATIENT INVOLVEMENT.