inforMED
MalfunctionFAS

HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°

Received Apr 3, 2026 · Event occurred Mar 6, 2026

Report 3007960282-2026-00059 · MDR key 24780652

Device

Generic name

Hf-resection Electrode

Model number

WA22306D

Lot number

1000194960

Product problems

  • Material Deformation

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: THE MOST PROBABLE CAUSE FOR THE MALFUNCTION WAS DUE TO COMPONENT FAILURE. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem

NO ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER.

Additional Manufacturer Narrative

E1: (B)(6). THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

IT WAS REPORTED THE RESECTION ELECTRODE HAD BROKE AT THE MOMENT OF OUTPUT. THE ISSUE WAS FOUND DURING A THERAPEUTIC TRANSURETHRAL RESECTION OF BLADDER TUMOR (TURBT) PROCEDURE AND COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.