ALINITY I FREE T4 REAGENT KIT
Received Apr 3, 2026 · Event occurred Feb 27, 2026
Report 3005094123-2026-00157 · MDR key 24777815
Device
Generic name
Radioimmunoassay, Free Thyroxine
Manufacturer
Abbott Ireland Diagnostics DivisionCatalog number
07P70-30
Lot number
79271UD00
Product problems
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE 9 - 23 PMOL/L): (B)(6) 2026, SID (B)(6), INITIAL FREE T4 RESULT= 48.6 PMOL/L; REPEAT RESULTS= 17.4 AND 16.6 PMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Additional Manufacturer Narrative
COMPLETE ENTRY FOR SECTION A1 - PATIENT IDENTIFIER: (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION.