inforMED

O.R. SOLUTIONS

Received Feb 29, 2012 · Event occurred Feb 22, 2012

Report 2476084 · MDR key 2476084

Device

Generic name

Warmer Drape

Manufacturer

O.r. Solutions

Model number

ORS-100

Catalog number

ORS-100

Lot number

0811003

Product problems

  • Hole In Material
  • Melted

Patient

42 YR

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

A HOLE WAS FOUND IN A STERILE IRRIGATION DRAPE. IT APPEARS TO HAVE MELTED. TESTING OF THREE SEPARATE WARMERS CONFIRMED THE TEMPERATURE WAS STABLE AND WITHIN TOLERANCE.======================MANUFACTURER RESPONSE FOR WARMER DRAPE, O.R. SOLUTIONS (PER SITE REPORTER).======================MANUFACTURER PROVIDED THE RETURN GOODS AUTHORIZATION (RGA) NUMBER FOR PRODUCT RETURN ANALYSIS.======================MANUFACTURER RESPONSE FOR WARMING UNIT, O.R. SOLUTIONS (PER SITE REPORTER).======================MANUFACTURER PROVIDED TESTING GUIDELINES TO TEST THE THREE UNITS THAT THE DRAPE COULD HAVE POSSIBLY BEEN USED. NO TESTING DISCREPANCIES WERE NOTED. ALL UNITS WERE WITHIN TOLERANCE. THE TEMPERATURE WAS 118-122 DEGREES FAHRENHEIT.