inforMED
MalfunctionFEM

WM-DP2 WORKSTATION SET 1 - UK

Received Apr 1, 2026 · Event occurred Mar 2, 2026

Report 9611174-2026-00071 · MDR key 24760497

Device

Generic name

Wm-*p2 Series Of Endoscopy Workstations

Model number

K10020641

Product problems

  • Smoking

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: SCORCHING ON BASE AND MELTED PLASTIC. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: SCORCHING ON BASE CAUSED BY A POWER CABLE SITTING ON THE BASE NOT PLUGGED IN. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A LARGE SPARK FROM THE UNDERNEATH SECTION OF THE ENDOSCOPY WORKSTATIONS WHERE ALL THE CABLES WERE HOUSED AND SOME SCORCHING. THE ISSUE WAS FOUND DURING SEDATION - DEVICE INSIDE PATIENT FOR AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE AND IT WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.