WM-DP2 WORKSTATION SET 1 - UK
Received Apr 1, 2026 · Event occurred Mar 2, 2026
Report 9611174-2026-00071 · MDR key 24760497
Device
Generic name
Wm-*p2 Series Of Endoscopy Workstations
Manufacturer
Keymed (medical And Industrial Equipment) Ltd.Model number
K10020641Product problems
- Smoking
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: SCORCHING ON BASE AND MELTED PLASTIC. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: SCORCHING ON BASE CAUSED BY A POWER CABLE SITTING ON THE BASE NOT PLUGGED IN. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Description of Event or Problem
IT WAS REPORTED THAT THERE WAS A LARGE SPARK FROM THE UNDERNEATH SECTION OF THE ENDOSCOPY WORKSTATIONS WHERE ALL THE CABLES WERE HOUSED AND SOME SCORCHING. THE ISSUE WAS FOUND DURING SEDATION - DEVICE INSIDE PATIENT FOR AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE AND IT WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.