Received Apr 1, 2026 · Event occurred Mar 9, 2026

Report 1823260-2026-01219 · MDR key 24757233

Device

Generic name

Radioimmunoassay, Estradiol

Catalog number

09629645190

Lot number

930252

Product problems

  • High Test Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE ROOT CAUSE WAS DETERMINED TO BE DUE TO CONTAMINATION WITH THE ANALYTE FROM THE ENVIRONMENT. PER THE PRODUCT LABELING, "ESTRADIOL IS SHED NATURALLY IN SKIN PARTICLES, SALIVA, AND OTHER BODY FLUIDS AND CAN EASILY BE DISTRIBUTED VIA DUST OR AEROSOLS.19,20 AS A RESULT, LAB ENVIRONMENTS INCLUDING SAMPLES, CONSUMABLES, OR INSTRUMENTS MAY BE CONTAMINATED WITH ESTRADIOL. HIGH SENSITIVE ESTRADIOL ASSAYS MAY BE IMPACTED BY FALSELY ELEVATED ESTRADIOL ASSAY VALUES. FOR DIAGNOSTIC PURPOSES, THE RESULTS SHOULD ALWAYS BE ASSESSED IN CONJUNCTION WITH THE PATIENT¿S MEDICAL HISTORY, CLINICAL EXAMINATION AND OTHER FINDINGS." THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.

Additional Manufacturer Narrative

THE SERIAL NUMBER OF THE ANALYZER IS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE ESTRADIOL RESULTS WITH THE IONIFY STEROIDS 1 ASSAY FOR SEVERAL PATIENT SAMPLES TESTED ON THE COBAS I 601 INSTRUMENT. SAMPLE 1: THE INITIAL RESULT WAS 1.50 PG/ML (ACCOMPANIED BY A DATA FLAG). ON (B)(6) 2026, THE SAMPLE WAS REPEATED TWICE AND THE RESULTS WERE 1.50 PG/ML (ACCOMPANIED BY A DATA FLAG) AND 13.8 PG/ML. SAMPLE 2: THE INITIAL RESULT WAS 80.4 PG/ML (ACCOMPANIED BY A DATA FLAG). THE SAMPLE WAS REPEATED TWICE AND THE RESULTS WERE 1.50 PG/ML (ACCOMPANIED BY A DATA FLAG) AND 2.42 PG/ML.