HYDROS ROBOTIC SYSTEM
Received Apr 1, 2026 · Event occurred Mar 16, 2026
Report 3012977056-2026-00065 · MDR key 24755952
Device
Product problems
- Adverse Event Without Identified Device or Use Problem
Patient
NA · Male
- Perforation
Narrative
Description of Event or Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS (PROCEPT) BECAME AWARE THAT DURING FOCAL BLADDER NECK CAUTERY (FBNC), THE SURGEON RESECTED OVER THE PATIENT'S RIGHT URETERAL ORIFICE (UO). IT IS UNKNOWN WHETHER IF THERE WERE ANY MEDICAL INTERVENTION OR ADDITIONAL SURGERY. ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION HOWEVER INFORMATION COULD NOT BE RETRIEVED REGARDING THE EVENT AND PATIENT STATUS. NO MALFUNCTION OF THE HYDROS ROBOTIC SYSTEM WAS REPORTED.
Additional Manufacturer Narrative
THE HYDROS ROBOTIC SYSTEM IS A REUSABLE DEVICE; THEREFORE, IT IS STILL CURRENTLY IN POSSESSION OF THE USER FACILITY. THE INVESTIGATION OF THIS EVENT CONSISTED OF A REVIEW OF THE TREATMENT LOG FILES, DEVICE HISTORY RECORD (DHR), INSTRUCTIONS FOR USE (IFU), AND PROCEPT'S RISK DOCUMENTATION. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR HY1000/SERIAL NUMBER: (B)(6) WAS CONDUCTED, WHICH CONFIRMED THAT THERE WERE NO NON-CONFORMANCE'S, FAILURES, DISCREPANCIES, OR MISSED STEPS DURING THE MANUFACTURING PROCESS THAT COULD BE RELATED TO THE REPORTED EVENT. THE REVIEW INDICATED THAT THE DEVICE MET ALL DESIGN AND MANUFACTURING SPECIFICATIONS WHEN RELEASED FOR DISTRIBUTION. THE HYDROS ROBOTIC SYSTEM'S TREATMENT LOG FILES WERE REVIEWED. THE REVIEW OF THE TREATMENT LOG FILES REVEALED THAT THE HYDROS ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AS NO MALFUNCTION WAS OBSERVED DURING AQUABLATION THERAPY. THE HYDROS ROBOTIC SYSTEM'S USER MANUAL (UM), UM0401-00-01, REV. C, WAS REVIEWED. ALTHOUGH THE HYDROS ROBOTIC SYSTEM'S LABELING DOES NOT SPECIFICALLY MENTION DAMAGE TO URETERAL ORIFICES, PROCEPT'S RISK MANAGEMENT DOCUMENTATION INCLUDES DAMAGE TO URETERAL ORIFICES AS A CLINICAL EFFECT OF AQUABLATION THERAPY. THE HYDROS ROBOTIC SYSTEM IS A REUSABLE DEVICE; THEREFORE, IT IS STILL CURRENTLY IN POSSESSION OF THE USER FACILITY. IT WAS REPORTED THAT DURING FOCAL BLADDER NECK CAUTERY (FBNC), THE SURGEON RESECTED OVER THE PATIENT'S RIGHT URETERAL ORIFICE (UO). IT IS UNKNOWN WHETHER IF THERE WERE ANY MEDICAL INTERVENTION OR ADDITIONAL SURGERY. ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION HOWEVER INFORMATION COULD NOT BE RETRIEVED REGARDING THE EVENT AND PATIENT STATUS. ALTHOUGH THE HYDROS ROBOTIC SYSTEM'S LABELING DOES NOT SPECIFICALLY MENTION DAMAGE TO URETERAL ORIFICES, PROCEPT'S RISK MANAGEMENT DOCUMENTATION INCLUDES DAMAGE TO URETERAL ORIFICES AS A CLINICAL EFFECT OF AQUABLATION THERAPY. BASED ON THE INFORMATION RECEIVED, PLUS A REVIEW OF THE TREATMENT LOG FILES, DHR, IFU, AND RISK DOCUMENTATION, THE EVENT IS CONSIDERED NOT TO BE DEVICE RELATED. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.