HEMOSIL SYNTHASIL

Received Apr 1, 2026 · Event occurred Mar 4, 2026

Report 1217183-2026-00002 · MDR key 24754148

Device

Generic name

Activated Partial Thromboplastin

Catalog number

00020006800

Lot number

N1046401

Product problems

  • Device Ingredient or Reagent Problem
  • Installation-Related Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

REVIEW OF THE INSTRUMENT SERVICE HISTORY INDICATED THAT THE CUSTOMER REPORTED INCORRECT LOW APTT RESULTS FOR SIX PATIENT SAMPLES, WITH REPEAT TESTING PRODUCING HIGHER VALUES. DURING THE ON-SITE SERVICE VISIT, THE FIELD SERVICE ENGINEER (FSE) REPORTED CUVETTE SHUTTLE MOVEMENT FAILURE AND UPON TROUBLESHOOTING, FOUND CUVETTE WAS UPSIDE DOWN INSIDE THE INSTRUMENT. ALSO, FSE CONFIRMED LOW APTT QUALITY CONTROL (QC) RESULTS. REVIEW OF THE INSTRUMENT ERROR LOGS DID NOT IDENTIFY ANY ERRORS RELATED TO THE REPORTED CONDITION. DIAGNOSTIC EVALUATION REVEALED THAT THE RED LED DIGITAL-TO-ANALOG CONVERTER (DAC) VALUE WAS 215, APPROACHING THE UPPER LIMIT OF THE OPERATING RANGE. THE RED LED WAS REPLACED AND THE DAC VALUES WERE ADJUSTED TO FALL WITHIN SPECIFICATION. FURTHER INSPECTION OF ON-BOARD REAGENTS IDENTIFIED FLOATING CONTAMINANTS IN THE FACTOR DILUENT BOTTLE. THE FSE PERFORMED AIR AND FACTOR DILUENT BLANKING, INSPECTED PROBES AND REAGENT AND SAMPLE FLUIDIC PATHWAYS, AND CONDUCTED FLUIDIC PRECISION AND FLOW RATE TESTING; ALL TESTS MET ACCEPTANCE CRITERIA. SUBSEQUENT QC AND PATIENT SAMPLE TESTS PERFORMED BY THE CUSTOMER YIELDED PASSING RESULTS, INDICATING RESOLUTION OF THE REPORTED ISSUE. NO PATIENT IMPACT REPORTED. BASED ON THIS ASSESSMENT, NO REMEDIAL ACTION IS REQUIRED. THIS INCIDENT WILL BE MONITORED THROUGH TRENDING ANALYSIS.

Description of Event or Problem

ON (B)(6) 2026, SIX PATIENT SAMPLES GENERATED ERRONEOUSLY LOW APTT RESULTS WHEN TESTED WITH HEMOSIL SYNTHASIL LOT N1046401 ON AN ACL TOP 300 CTS (SN (B)(6)). THE SAMPLES WERE RERUN AND GENERATED HIGHER RESULTS, AFTER WHICH CORRECTED PATIENT REPORTS WERE ISSUED. DURING TROUBLESHOOTING FOR CUVETTE SHUTTLE MOVEMENT ERRORS, IT WAS DETERMINED THAT CUVETTES HAD BEEN LOADED UPSIDE DOWN. QUALITY CONTROL AUTO-EXECUTED AND FAILED LOW FOR BOTH LEVELS. SOON AFTER, THE APTT-SS REAGENT GENERATED A "LEVEL LOW" WARNING; THE REAGENT WAS REPLACED AND QC PASSED WITHIN THE EXPECTED RANGE. SERVICE WAS DISPATCHED TO THE SITE. NO PATIENT IMPACT WAS REPORTED.