ACHIEVA 3.0 T
Received Apr 1, 2026 · Event occurred Mar 24, 2026
Report 3042177665-2026-000020 · MDR key 24751012
Device
Generic name
Full-body Mri System, Superconducting Magnet
Manufacturer
Philips Medical Systems Nederland B.v.Model number
ACHIEVA 3.0 TCatalog number
781177
Product problems
- Detachment of Device or Device Component
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE QBC SEAL THAT HAD COME LOOSE WAS REPLACED, AND THE SYSTEM WAS RETURNED TO THE CUSTOMER FOR USE. THE FAILURE OF THE QBC SEAL TO REMAIN IN PLACE IS CONSIDERED A MALFUNCTION. REMEDIAL ACTION: A FIELD ACTION WAS INITIATED UNDER 2023-PD-MR-013. A FIELD SAFETY NOTIFICATION WAS SENT TO THE CUSTOMERS INFORMING THEM ABOUT THIS POTENTIAL ISSUE. THE FIELD CORRECTION HAS SINCE BEEN MADE AVAILABLE TO THE FIELD IN DEC 2024.
Description of Event or Problem
PHILIPS RECEIVED A REPORT THAT THE RUBBER PART THAT CONNECTED THE BODY COIL TO THE PLASTIC INLET HAD COME LOOSE. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE, IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.
Remedial action
- Modification/Adjustment