inforMED
InjuryKWI

EXPLOR 12X20MM IMPLANT HEAD

Received Mar 30, 2026 · Event occurred Jun 24, 2019

Report 0001825034-2026-00749 · MDR key 24716380

Device

Generic name

Elbow Radius Prosthesis

Catalog number

11-210023

Lot number

424780

Product problems

  • Inadequacy of Device Shape and/or Size

Patient

NA · Male

  • Extraskeletal ossification
  • Pain
  • Loss of Range of Motion
  • Scar Tissue
  • Joint Contracture
  • Aseptic loosening

Narrative

Additional Manufacturer Narrative

(B)(4). D10: ITEM #: 11-210063, EXPLOR 8X28MM IMPL STEM W/SCR, LOT # 811940. G2: FOREIGN - EVENT OCCURRED IN NORWAY. H6: PROPOSED COMPONENT CODE - MECHANICAL (G04) - HEAD. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED THAT AFTER 21 DAYS POST IMPLANTATION PATIENT PRESENTED WITH CONTRACTURE ASSOCIATED WITH LOSS OF ROTATION, FLEXION, AND EXTENSION (0°, 100°, 20°). X-RAYS SHOWED EVIDENCE OF ECTOPIC BONE FORMATION. INTRAOPERATIVELY, SIGNS OF JOINT ¿STUFFING¿ AND A HEAD SIZE CONSIDERED TOO LARGE WERE OBSERVED. AN OPEN ARTHROLYSIS WAS PERFORMED, INCLUDING REMOVAL OF SCAR TISSUE AND ECTOPIC BONE FORMATION. SCARRING AGAINST THE TROCHLEA FROM THE PROSTHESIS HEAD WAS ALSO IDENTIFIED. THE PROSTHETIC HEAD WAS REMOVED AND REPLACED WITH A SMALLER HEAD MEASURING 20 MM WITH A 10 MM LENGTH. POST-OPERATIVE RADIATION THERAPY WAS PLANNED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.