inforMED
MalfunctionFMZ

GIRAFFE OMNIBED CARESTATION

Received Mar 30, 2026 · Event occurred Feb 27, 2026

Report 3008729547-2026-00027 · MDR key 24713713

Add to watch list
Product code (FMZ)
Manufacturer (Ge Medical Systems Information Technologies - Critikon De Mexico S. De R.l. De C.v.)
Model (CS1)

Device

Generic name

Incubator, Neonatal

Model number

CS1

Catalog number

NA

Lot number

NA

Product problems

  • No Flow

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

GE HEALTHCARE¿S INVESTIGATION INTO THE REPORTED OCCURRENCE IS ONGOING. A FOLLOW-UP REPORT WILL BE ISSUED WHEN THE INVESTIGATION HAS BEEN COMPLETED. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS RESEARCH PARK 9900 INNOVATION DRIVE USA WAUWATOSA, WI 53226.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN OXYGEN FAILURE DURING SERVICE. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE SERVO O2 BOARD WAS REPLACED TO RESOLVE THE ISSUE.