inforMED
MalfunctionCBL

LITHOLYME

Received Mar 26, 2026 · Event occurred Jan 29, 2026

Report 1924066-2026-00004 · MDR key 24707078

Device

Generic name

Litholyme, Ge Multi-style Cartridge

Model number

55-01-0017

Catalog number

55-01-0017

Lot number

N/A

Product problems

  • Gas/Air Leak

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

NO PATIENT HARM OCCURRED; HOWEVER, THE EVENT REQUIRED IMMEDIATE INTERVENTION TO MAINTAIN ADEQUATE VENTILATION. THE DEVICE WAS NOT RETURNED FOR INVESTIGATION, PREVENTING FURTHER ANALYSIS OF THE CAUSE.

Description of Event or Problem

WHEN CHANGING THE CO2 ABSORBENT, IT BECOMES DIFFICULT TO SEAL. RESULTS IN A CIRCUIT LEAK. VERY DANGEROUS TO PATIENT. FLOW SENSORS O2 RINGS FELL OUT OF MACHINE ALONG WITH THE ABSORBENT CONTAINERS, CAUSING CRNA TO MANUALLY VENTILATE THE PATIENT UNTIL THE ANES. TECH WAS ABLE TO PUT PART BACK IN.