#23/UNC12 EXPRO #7 HANDLE
Received Mar 26, 2026 · Event occurred Mar 5, 2026
Report 1416605-2026-00004 · MDR key 24702145
Device
Generic name
Expro Probe
Manufacturer
Hu-friedy Manufacturing Co., LlcModel number
XP23/NC127Catalog number
XP23/NC127
Lot number
0425
Product problems
- Break
Patient
NA · Unknown
- Foreign Body In Patient
Narrative
Description of Event or Problem
THE USER FACILITY REPORTED THAT DURING A DENTAL PROCEDURE THE TIP OF THE PROBE BROKE IN PATIENT'S MOUTH AND THE PATIENT MIGHT HAVE SWALLOWED IT. A RADIOGRAPH WAS TAKEN TO LOCATE THE TIP AND THEY FOUND NO EVIDENCE OF THE TIP. PATIENT DECLINED ANY FURTHER MEDICAL INTERVENTION.
Additional Manufacturer Narrative
THE DEVICE SUBJECT OF THE REPORTED EVENT WAS RETURNED TO HU-FRIEDY FOR EVALUATION. THE PROBE TIP WAS LIKELY SUBJECTED TO EXCESSIVE FORCE DURING USE, LEADING TO THE BREAKAGE OF THE POINT. THIS INDICATES A USE-RELATED FAILURE MECHANISM. THE DEVICE HISTORY RECORD FOR THE SUBJECT LOT WAS REVIEWED. RETURN RATES AND COMPLAINT RATES ARE LOW, NO FURTHER ACTION WILL BE TAKEN.