inforMED
MalfunctionDXO

TRUWAVE DISPOSABLE PRESSURE TRANSDUCER

Received Mar 26, 2026 · Event occurred Mar 10, 2026

Report 2015691-2026-13553 · MDR key 24698083

Device

Generic name

Transducer, Pressure, Catheter Tip

Model number

PX260

Lot number

65944748

Product problems

  • Contamination

Patient

76 YR · Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS; HOWEVER, IT HAS NOT YET BEEN RECEIVED. UPON THE RETURN OF THE PRODUCT A SUPPLEMENTAL REPORT WILL BE SENT WITH THE INVESTIGATION RESULTS. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW.

Description of Event or Problem

AS REPORTED, DURING USE IN PATIENT, INSPECTION OF THIS DISPOSABLE PRESSURE TRANSDUCER (DPT) REVEALED THAT THE FEMALE LUER ON THE DPT HOUSING, WHICH CONNECTS TO THE IV TUBING, WAS BLOCKED BY A FOREIGN MATERIAL. THE PROBLEM WAS SOLVED BY REPLACING THE DEVICE WITH A NEW ONE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION; HOWEVER, IT HAS NOT BEEN RECEIVED YET.