TRUWAVE DISPOSABLE PRESSURE TRANSDUCER
Received Mar 26, 2026 · Event occurred Mar 11, 2026
Report 2015691-2026-13552 · MDR key 24698026
Device
Generic name
Transducer, Pressure, Catheter Tip
Manufacturer
Edwards Lifesciences DrModel number
PX260Lot number
65944748
Product problems
- Low Readings
Patient
65 YR · Male
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
AS REPORTED, AFTER CONNECTING THIS PRESSURE MONITORING SET TO THE PATIENT'S RADIAL ARTERY, THE PRESSURE VALUES OBTAINED (75/47 MMHG) DIFFERED SIGNIFICANTLY FROM THE PREOPERATIVE NON-INVASIVE CUFF MEASUREMENT (135/85 MMHG). NO ABNORMAL ALARMS OR ALERTS WERE OBSERVED. THE PRESSURE MONITORING SET WAS REPLACED WITH A NEW UNIT, AFTER WHICH NORMAL PRESSURE READINGS (143/88 MMHG) WERE OBTAINED. THE PATIENT DID NOT RECEIVE ANY TREATMENT BASED ON THE INACCURATE READINGS. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION; HOWEVER, IT HAS NOT BEEN RECEIVED YET.
Additional Manufacturer Narrative
THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS; HOWEVER, IT HAS NOT YET BEEN RECEIVED. UPON THE RETURN OF THE PRODUCT A SUPPLEMENTAL REPORT WILL BE SENT WITH THE INVESTIGATION RESULTS. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW.