I-STAT CHEM8+ CARTRIDGE
Received Mar 26, 2026 · Event occurred Feb 17, 2026
Report 2245578-2026-00072 · MDR key 24697515
Device
Generic name
Chem8+ Cartridge
Manufacturer
Abbott Point Of CareModel number
NACatalog number
09P31-26
Lot number
H25310
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
NA · Female
- Muscle Weakness/Atrophy
- Facial Paralysis
Narrative
Additional Manufacturer Narrative
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 14-APR-2026. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS RELEASE CRITERIA. RETAINED AND RETURNED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AP, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE, EXCEPT FOR THE RATE OF IONIZED CALCIUM RESULTS OUTSIDE OF THE TOTAL ALLOWABLE ERROR WHEN TESTING IS PERFORMED USING WHOLE BLOOD (WB). A PREVIOUSLY IDENTIFIED DEFICIENCY FOR ICA RESULTS OUTSIDE OF EA, WHICH IS UNRELATED TO THE CUSTOMER'S OBSERVATION OF NA AND CREA DISCREPANT RESULTS, HAS BEEN DETERMINED FOR CHEM8+ CARTRIDGE LOT H25310 AND IS UNDER INVESTIGATION IN QUALITY RECORD 1096422.
Description of Event or Problem
NA.
Additional Manufacturer Narrative
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
Description of Event or Problem
ON (B)(6) 2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED A DISCREPANT SODIUM RESULT ON A 25-YEAR-OLD MALE PATIENT WITH RIGHT-SIDED WEAKNESS, FACIAL DROOP. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD DATE RESULT I-STAT (B)(6) 2026 166 LAB (B)(6) 2026 138 I-STAT (B)(6) 2026 140. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.