inforMED
MalfunctionOAX

BINAXNOW MALARIA 25T US

Received Mar 25, 2026 · Event occurred Feb 24, 2026

Report 1221359-2026-00038 · MDR key 24688109

Device

Generic name

Plasmodium Spp. Detection Reagents

Catalog number

665-025

Lot number

0000914228

Product problems

  • False Negative Result

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER REPORTED THREE (3) FALSE NEGATIVE RESULTS OBTAINED WITH BINAXNOW MALARIA TESTS ON A SINGLE PATIENT ACROSS MULTIPLE TESTING DATES. THIS MANUFACTURER REPORT PERTAINS TO TEST THREE (3) OF THE THREE (3) TESTS. THE CUSTOMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW MALARIA TEST PERFORMED ON (B)(6) 2026 ON A WHOLE BLOOD SAMPLE. SUBSEQUENT CONFIRMATION TESTING WAS CONDUCTED VIA MICROSCOPY AND PCR ON WHOLE BLOOD SAMPLES ON (B)(6)2026 AND (B)(6)2026, BOTH OF WHICH YIELDED POSITIVE RESULTS. ACCORDING TO THE CUSTOMER, THE DELAYED DETECTION OF MALARIA RESULTED IN DELAYED IDENTIFICATION OF THE INFECTION, AND TREATMENT FOR A POSSIBLE PLASMODIUM OVALE/VIVAX INFECTION WAS INITIATED.

Additional Manufacturer Narrative

THE REMAINDER OF THE INVESTIGATION IS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED PENDING COMPLETION, OR UPON RECEIPT OF NEW INFORMATION.