inforMED
MalfunctionFMT

PANDA

Received Mar 25, 2026 · Event occurred Feb 25, 2026

Report 3036470679-2026-00034 · MDR key 24687682

Device

Generic name

Warmer, Infant Radiant

Manufacturer

Datex-ohmeda, Inc.

Model number

IRES WARMER

Catalog number

NA

Lot number

NA

Product problems

  • No Flow

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE O2 REGULATOR WAS REPLACED TO RESOLVE THE ISSUE.

Additional Manufacturer Narrative

GE HEALTHCARE¿S INVESTIGATION INTO THE REPORTED OCCURRENCE IS ONGOING. A FOLLOW-UP REPORT WILL BE ISSUED WHEN THE INVESTIGATION HAS BEEN COMPLETED. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: GEHC TROUT - 3114 N GRANDVIEW BLVD. USA WAUKESHA, WI 53188.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF OXYGEN THERAPY. THERE WAS NO PATIENT INVOLVEMENT REPORTED.