Generic name

Intramedullary Fixation Rod

Catalog number

PPFR11420

Lot number

X2647

Product problems

  • Fracture

Patient

NA · Male

  • Bone Fracture(s)
  • Nonunion/Delayed-union Bone Fracture

Narrative

Additional Manufacturer Narrative

EXAMINATION OF THE PRODUCTION RECORDS OF THE INVOLVED NAIL DID NOT REVEAL ANY DEVIATION FROM SPECIFICATION. THE EXPLANTED NAIL (NOR THE CASE X-RAYS) WAS NOT AVAILABLE TO THE MANUFACTURER; THEREFORE, NAIL PHYSICAL EXAMINATION WAS NOT FEASIBLE. ALTHOUGH THE NAIL COULD NOT BE EXAMINED, THE FACT THAT IT FRACTURED AFTER FIVE MONTHS IN METASTATIC BONE WITH NON-UNION SUGGESTS A FATIGUE FRACTURE. BREAKAGE OF PROXIMAL FEMUR NAILS IS KNOWN AND DOCUMENTED IN THE LITERATURE, WITH REPORTED RATES THAT RANGE FROM 0.2% TO 5.6%. THE RISK OF IMPLANT FAILURE INCREASES IN CASE OF PATHOLOGICAL FRACTURES, AS IN THE DESCRIBED CASE. [1] THE COMPANY IS ACTIVELY SEEKING FOR MORE DATA. [1] JOHNSON NA ET AL., RISK FACTORS FOR INTRAMEDULLARY NAIL BREAKAGE IN PROXIMAL FEMORAL FRACTURES: A 10-YEAR RETROSPECTIVE REVIEW. ANN R COLL SURG ENGL. 2017; 99(2): 145-150.

Description of Event or Problem

ON MARCH 16, 2026, THE COMPANY BECAME AWARE OF A CASE INVOLVING ITS PICCOLO COMPOSITE PF NAIL. BASED ON THE INFORMATION PROVIDED TO THE COMPANY, THE DEVICE WAS IMPLANTED IN AN ONCOLOGIC PATIENT WITH FEMORAL METASTASES ON (B)(6) 2024, FOLLOWING A FALL RESULTING IN A PATHOLOGICAL FRACTURE OF THE FEMUR. APPROXIMATELY FIVE MONTHS POST-IMPLANTATION, THE NAIL FRACTURED AT OR NEAR THE LAG SCREW INTERFACE DUE TO NON-UNION. ON (B)(6) 2025, THE NAIL WAS REMOVED AND REPLACED WITH A TOTAL HIP PROSTHESIS.