inforMED
MalfunctionDTB

RETROX RX 45-JBP

Received Mar 23, 2026

Report 1028232-2026-01432 · MDR key 24664537

Device

Generic name

Permanent Pacemaker Electrode

Model number

124395

Catalog number

124395

Product problems

  • Failure to Capture
  • Over-Sensing
  • High Capture Threshold

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE IS CURRENTLY NOT AVAILABLE FOR ANALYSIS. NO CONCLUSION CAN BE DRAWN AT THIS TIME. NO ADDITIONAL INFORMATION IS AVAILABLE AT THE MOMENT. THE FILE IS CLOSED. THE INVESTIGATION WILL BE RE-OPENED SHOULD ADDITIONAL DATA BECOME AVAILABLE.

Description of Event or Problem

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVER SENSING, HIGH THRESHOLDS AND NO CAPTURE. DISCLAIMER STATEMENT: THIS REPORT REFLECTS INFORMATION RECEIVED BY FDA IN THE FORM OF A NOTIFICATION PER 803.22 (B)(2). MW5183491.