inforMED
InjuryOOE

LENSAR LASER SYSTEM-FS 3D

Received Mar 23, 2026 · Event occurred Mar 9, 2026

Report 3009026057-2026-00013 · MDR key 24664129

Device

Generic name

Lensar Laser System-fs 3d

Manufacturer

Lensar, Inc.

Model number

N/A

Catalog number

70-00005-001

Lot number

N/A

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

NA · Unknown

  • Capsular Bag Tear

Narrative

Description of Event or Problem

ON (B)(6) 2026, DOCTOR ((B)(6)) REPORTED TO CAS THAT HE HAD AN ANTERIOR CAPSULE RUN OUT AT THE BASE OF THE TORIC NUB DURING PROCEDURE ID (B)(6) ON (B)(6) 2026. SITE IS SEEKING REVIEW OF THE PROCEDURE.

Additional Manufacturer Narrative

REVIEW OF PROCEDURE FILES AND IMAGING SHOW THE CAPSULOTOMY STARTS ON FRAME 3 AS EXPECTED AND THE CAPSULOTOMY STARTS TO BREAK THROUGH INTO THE AC ON FRAME 5 AND COMPLETELY THROUGH ON FRAME 7.THERE IS NOTHING NOTEWORTHY ON THIS CASE. SYSTEM FUNCTIONED AS DESIGNED. NO FURTHER CLINICAL FOLLOW-UP REQUIRED.