BD VERITOR¿ PLUS ANALYZER

Received Mar 20, 2026 · Event occurred Feb 12, 2026

Report 1119779-2026-00441 · MDR key 24656487

Device

Generic name

Colorimeter, Photometer, Spectrophotometer For Clinical Use

Catalog number

256066

Product problems

  • False Negative Result

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

REPORT 1 OF 2 IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, ONE (1) NEGATIVE PATIENT RESULT WAS OBTAINED. THE SAME TEST CARTRIDGE WAS REINSERTED INTO A DIFFERENT VERITOR ANALYZER WHICH PROVIDED A POSITIVE RESULT. IN ADDITION, THE PATIENT SAMPLE WAS SENT TO A DIFFERENT LAB AND CONFIRMED TO BE POSITIVE BY AN UNKNOWN TEST METHODOLOGY. THE CUSTOMER IS ABLE TO PROVIDE THAT THE ASSAY TESTED WAS INFLUENZA RAPID DETECTION OF FLU A AND B; HOWEVER, A CATALOG NUMBER IS UNABLE TO BE PROVIDED. IT IS NOTED THAT AN ADDITIONAL SWAB WAS NOT COLLECTED FOR RETESTING AND THE CUSTOMER DID NOT SPECIFY IF THE ANALYTE IN QUESTION WAS FLU A OR FLU B. IT SHOULD ALSO BE NOTED THAT REINSERTING A USED TEST CARTRIDGE IS CONSIDERED OFF-LABEL USE OF THE PRODUCT. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

Additional Manufacturer Narrative

THE REQUESTED 510K INFORMATION FOR THE INVOLVED VERITOR ASSAY IS AS FOLLOWS: G4. PMA/510(K)#: CUSTOMER COULD NOT PROVIDE THE ASSAY MEDICAL DEVICE CATALOG # TO DETERMINE THE 510K. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.