MalfunctionMLR
V-PRO
Received Mar 19, 2026 · Event occurred Feb 14, 2026
Report 3005899764-2026-00024 · MDR key 24643198
Device
Generic name
Sterilizer
Manufacturer
Steris Mexico, S. De R.l. De C.v.Model number
MAX 2Catalog number
VP50002101
Product problems
- Leak/Splash
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE USER FACILITY REPORTED AN OIL LEAK FROM THEIR V-PRO MAX 2 STERILIZER. NO REPORT OF INJURY.
Additional Manufacturer Narrative
A STERIS SERVICE TECHNICIAN ARRIVED ON-SITE FOLLOWING THE REPORTED EVENT TO INSPECT THE V-PRO MAX 2 STERILIZER AND FOUND THAT THE UNIT'S VACUUM PUMP SEAL WAS DAMAGED AND LEAKING OIL. THE TECHNICIAN REPLACED THE VACUUM PUMP AND SEAL, TESTED THE UNIT, CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED IT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.