inforMED
MalfunctionDTS

MEDTRONIC MINI SUCKER

Received Mar 19, 2026 · Event occurred Mar 17, 2026

Report 3003955307-2026-00020 · MDR key 24638017

Device

Generic name

Sucker, Cardiotomy Return, Cardiopulmonary Byp

Model number

10062

Catalog number

10062

Lot number

0231912883

Product problems

  • Obstruction of Flow

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DEVICE EVALUATION SUMMARY: VISUAL INSPECTION OF THE UNOPENED DEVICE SHOWS EVIDENCE OF A BLOCKAGE AT THE TIP. A SYRINGE FILLED WITH DI WATER WAS CONNECTED TO THE DEVICE ALONG WITH A PRESSURE GAGE. WHEN THE SYRINGE WAS ENGAGED THERE WAS NO FLOW OBSERVED THROUGH THE DEVICE. REASON FOR THE RETURN WAS CONFIRMED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem

MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THE DLP RIGID SUCTION TUBE, IT WAS REPORTED THAT THE TIP WAS BLOCKED STRAIGHT OUT OF THE PACKAGE. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.