Device

Generic name

Ptx-iq System

Manufacturer

Litecure, Llc

Model number

SYS-000640-000-01

Product problems

  • Temperature Problem

Patient

NA · Unknown

  • Burn(s)

Narrative

Additional Manufacturer Narrative

CUSTOMER REPORTS LASER HAS BEEN BURNING PATIENTS RECENTLY. CUSTOMER TESTED IN ON THEIR SKIN AND THEY BURNED THEMSELVES - SAME THING HAPPENED WITH ANOTHER STAFF MEMBER. THE DEVICE HAS NOT BEEN RECEIVED FOR EVALUATION; THE COMPLAINT COULD NOT BE CONFIRMED; THE ROOT CAUSE COULD NOT BE ESTABLISHED. ADDITIONAL REPORTING ON THIS EVENT WILL BE PROVIDED AS A SUPPLEMENTAL REPORT TO THIS DOCUMENT IF NEW INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

CUSTOMER REPORTS LASER HAS BEEN BURNING PATIENTS RECENTLY. CUSTOMER TESTED IN ON THEIR SKIN AND THEY BURNED THEMSELVES - SAME THING HAPPENED WITH ANOTHER STAFF MEMBER.

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED AND EVALUATED BY A TECHNICIAN UNIT RAN MULTIPLE SETTINGS AND THE POWER OUTPUT IS WITHIN SPECIFICATIONS. THE UNIT IS NOT OVERPOWERING. A VERIFICATION FILE WAS RAN AND PASSED. THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE COULD NOT BE ESTABLISHED.