ELECSYS HE4

Received Mar 18, 2026 · Event occurred Feb 25, 2026

Report 1823260-2026-00983 · MDR key 24628416

Add to watch list
Product code (OIU)
Manufacturer (Roche Diagnostics International Ltd.)
Model (07027478190)

Device

Generic name

Human Epididymis Protein 4 (he4) Ivd, Kit, Chemiluminescent Immunoassay

Manufacturer

Roche Diagnostics

Catalog number

07027478190

Lot number

84981601

Product problems

  • High Test Results
  • Low Test Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE ANALYZER SERIAL NUMBER IS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS HE4 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS E 801 ANALYTICAL UNIT. THE INITIAL RESULT WAS 15.0 PMOL/L WITH FLAG. THE RESULT DID NOT MATCH THE PATIENT'S CLINICAL DIAGNOSIS WHICH PROMPTED THE CUSTOMER TO REPEAT THE SAMPLE. THE REPEAT RESULT WAS 73.9 PMOL/L. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY.

Additional Manufacturer Narrative

THE QC RECOVERY DATA PROVIDED WAS WITHIN SPECIFICATION. BASED ON THE QC DATA, A GENERAL REAGENT ISSUE COULD BE EXCLUDED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED.