AMSCO
Received Mar 18, 2026 · Event occurred Feb 17, 2026
Report 3005899764-2026-00022 · MDR key 24627720
Device
Generic name
Sterilizer
Manufacturer
Steris Mexico, S. De R.l. De C.v.Model number
AMSCO 600 STEAM STERILIZERCatalog number
EF1101142
Product problems
- Device Slipped
Patient
NA · Unknown
- Abrasion
Narrative
Additional Manufacturer Narrative
A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE UNIT AND FOUND THAT THE BRACKETS OF THE RACK AND SHELF ASSEMBLY WERE BENT CAUSING THE SHELVES TO SLIP OFF. TO RESOLVE THE ISSUE, THE TECHNICIAN ORDERED A NEW RACK AND SHELF ASSEMBLY. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE REPAIRS HAVE BEEN MADE. A 3-YEAR COMPLAINT REVIEW INDICATES THIS TO BE AN ISOLATED EVENT. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.
Description of Event or Problem
THE USER FACILITY REPORTED THAT WHILE AN EMPLOYEE WAS UNLOADING INSTRUMENTS FROM THEIR AMSCO 600 STERILIZER, A LOADED SHELF SLIPPED OFF BRACKET AND CONTACTED THE EMPLOYEE'S HAND RESULTING IN A BRUISE. NO MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED.
Additional Manufacturer Narrative
THE TECHNICIAN REPLACED THE RACK AND SHELF ASSEMBLY, TESTED THE UNIT, CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATIONS, AND RETURNED IT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.