inforMED
MalfunctionPZS

PROTEK D+ ECLS 31FR CAN

Received Mar 18, 2026 · Event occurred Feb 13, 2026

Report 1718850-2026-90003 · MDR key 24625437

Device

Generic name

Dual Lumen Ecmo Cannula

Catalog number

5150-5131

Lot number

UNKNOWN

Product problems

  • Fracture

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A.1.- A.5. PATIENT INFORMATION WAS NOT PROVIDED. D4. THE LOT NUMBER WAS NOT PROVIDED BY THE CUSTOMER. H11: THROUGH FOLLOW-UP COMMUNICATIONS, LIVANOVA LEARNED THAT THE CANNULA WAS USED FOR TWO (2) DAYS. AFTER THE CRACK WAS NOTICED, THE CANNULA WAS REPLACED WITH ANOTHER ONE. REPORTEDLY, THE CRACK DID NOT CAUSE ANY LEAKAGE OR FLOW ISSUE. IN ADDITION, THE CANNULA WAS SECURED WITHOUT A SUTURE RING. CUSTOMER REPORTEDLY DID NOT FIX THE PROXIMAL LEGS OF THE CANNULA TOGETHER DURING USE. LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem

LIVANOVA US RECEIVED REPORT OF A PROTEKDUO CANNULA WHICH CRACKED AT THE Y CONNECTOR DURING USAGE. NO PATIENT IMPACT WAS REPORTED.

Additional Manufacturer Narrative

G1: UPDATED G1 CONTACT OFFICE -(B)(6).

Description of Event or Problem

SEE INITIAL REPORT.