inforMED
MalfunctionFMZ

GIRAFFE INCUBATOR CARESTATION

Received Mar 17, 2026 · Event occurred Feb 17, 2026

Report 3008729547-2026-00024 · MDR key 24618374

Device

Generic name

Incubator, Neonatal

Model number

CS1

Catalog number

NA

Lot number

NA

Product problems

  • Insufficient Flow or Under Infusion

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

GE HEALTHCARE¿S INVESTIGATION INTO THE REPORTED OCCURRENCE IS ONGOING. A FOLLOW-UP REPORT WILL BE ISSUED WHEN THE INVESTIGATION HAS BEEN COMPLETED. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS RESEARCH PARK - 9900 INNOVATION DRIVE USA WAUWATOSA, WI 53226.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF OXYGEN THERAPY DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE OXYGEN SERVO BOARD WAS REPLACED TO RESOLVE THE ISSUE.