inforMED
MalfunctionDXO

PRESSUREWIRE¿

Received Mar 17, 2026 · Event occurred Oct 27, 2025

Report 2024168-2026-01051 · MDR key 24618010

Device

Generic name

Transducer, Pressure, Catheter Tip

Model number

C12059

Catalog number

C12059

Lot number

50619G1

Product problems

  • Break
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A VISUAL INSPECTION AND ADDITIONAL TESTING METHODS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED FAULTY PRESSURE REGISTRATION WAS NOT ABLE TO BE CONFIRMED. PRODUCTION RECORD AND CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT SPECIFIC ISSUE. BASED ON THE REPORTED INFORMATION AND THE OBSERVATIONS FROM THE RETURNED ANALYSIS, THE INVESTIGATION DETERMINED THAT THE REPORTED FAULTY PRESSURE REGISTRATION APPEARS TO BE RELATED TO OPERATIONAL CONTEXT. THE DEVICE WAS RETURNED WITH DAMAGE TO THE PRESSURE SENSOR OF THE GUIDEWIRE, WHICH RESULTED IN THE REPORTED AND CONFIRMED PRESSURE REGISTRATION ISSUE. THE PRESSURE SENSITIVITY OF THE GUIDEWIRE IS TESTED DURING THE MANUFACTURING PROCESS AND PRIOR TO RELEASE TO DEVICE FUNCTIONALITY UPON EXPECTED USE. IN THIS CASE, IT IS LIKELY THAT THE DAMAGE TO THE PRESSURE MEMBRANE OF THE SENSOR CHIP OCCURRED DUE TO THE HANDLING OR USE TECHNIQUES EMPLOYED. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.

Description of Event or Problem

REPORTEDLY, THE PWX DEVICE WAS CALIBRATED SUCCESSFULLY. WHILE ATTEMPTING EQUALIZATION, THERE WAS A LARGE DIFFERENCE BETWEEN THE PA AND PD VALUE, AND EQUALIZATION COULD NOT BE ACHIEVED. THE TRANSMITTER WAS GREEN. THEREFORE, THE DEVICE WAS REMOVED AND ANOTHER PWX DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED. DURING PRODUCT RETURN INVESTIGATION, IT WAS OBSERVED THAT THERE WAS A BREAK IN THE PRESSURE SENSOR MEMBRANE.