GIRAFFE OMNIBED CARESTATION
Received Mar 17, 2026 · Event occurred Feb 16, 2026
Report 3008729547-2026-00023 · MDR key 24615942
Device
Generic name
Incubator, Neonatal
Manufacturer
Critikon De Mexico S. De R.l. De C.v.Model number
CS1Catalog number
NA
Lot number
NA
Product problems
- No Flow
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
GE HEALTHCARE¿S INVESTIGATION INTO THE REPORTED OCCURRENCE IS ONGOING. A FOLLOW-UP REPORT WILL BE ISSUED WHEN THE INVESTIGATION HAS BEEN COMPLETED. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS RESEARCH PARK - 9900 INNOVATION DRIVE USA WAUWATOSA, WI 53226.
Description of Event or Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN OXYGEN FAILURE DURING SERVICE. THERE WAS NO PATIENT INVOLVEMENT.
Additional Manufacturer Narrative
A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE SUPPLY VALVE WAS REPLACED TO RESOLVE THE ISSUE.