MASSTRAK IMMUNOSUPPRESSANTS KIT
Received Dec 20, 2011 · Event occurred Nov 16, 2011
Report 1218959-2011-00002 · MDR key 2460622
Device
Generic name
Tacrolimus Test System
Manufacturer
Waters CorporationModel number
NACatalog number
186002891
Lot number
4110
Product problems
- Use of Device Problem
- Improper or Incorrect Procedure or Method
Patient
UNKNOWN
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
EVALUATION SUMMARY: REVIEW OF THE LABORATORY METHOD EMPLOYED BY (B)(6) REVEALED DEVIATIONS FROM THE WATERS MASSTRACK IMMUNOSUPPRESSANTS KIT - ANALYSIS OF TACROLIMUS DIRECTIONS FOR USE (DFU) INCLUDING USE OF AN INCORRECT SOLVENT TO PREPARE A SAMPLE PRETREATMENT REAGENT. THE MASSTRACK IMMUNOSUPPRESSANTS KIT FROM THE REPORTED LOT (#4110) WAS EVALUATED ON SITE ON (B)(4) 2011 BY A WATERS SERVICE ENGINEER USING THE WATERS ACQUITY TQD MASS SPECTROMETER (B)(4). TESTING WAS PERFORMED IN ACCORDANCE WITH THE DFU INCLUDING USE OF THE CORRECT SOLVENT TO PREPARE THE SAMPLE PRETREATMENT REAGENT. A CALIBRATION SERIES INCLUDING PT SAMPLES WITH KNOWN TACROLIMUS PROFILES WAS TESTED AND THE RESULTS OF THIS TEST ALIGNED WITH THE VALUES EXPECTED BY THE CUSTOMER. THIS CUSTOMER IS CURRENTLY PERFORMING THE ASSAY PER THE DFU AND HAS NOT SEEN A RECURRENCE OF THE REPORTED ISSUE. THE COMPLAINT REMAINS OPEN AS WE CONTINUE THE INVESTIGATION TO CONFIRM CAUSATION. SHOULD ADDITIONAL INFO BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE FILED.
Description of Event or Problem
A WATERS LABORATORY CUSTOMER REPORTED THAT SIXTEEN INACCURATELY HIGH TACROLIMUS VALUES WERE REPORTED TO PTS' PHYSICIANS ON (B)(6) 2011 AS A RESULT OF A THERAPEUTIC DRUG MONITORING (TDM) ASSAY RUN USING THE MASSTRAK IMMUNOSUPPRESSANTS KIT FOR ANALYSIS OF TACROLIMUS. THE LABORATORY REPEATED THE TEST AND NOTIFIED THE PHYSICIANS OF THE CORRECT RESULTS ON (B)(6) 2011. PER THE CUSTOMER, THIS INCIDENT RESULTED IN DOSE CHANGES FOR TWO PTS. THE CUSTOMER BELIEVES THE ISSUE IS DUE TO THE SAMPLE EXTRACTION (PRETREATMENT) METHOD THEY USED AND REPORTS THAT THEY DEVIATED FROM THE WATERS MASSTRAK IMMUNOSUPPRESSANTS KIT-ANALYSIS OF TACROLIMUS DIRECTIONS FOR USE (DFU) BY USING METHANOL SOLVENT RATHER THAN ACETONITRILE TO PREPARE A SAMPLE PRETREATMENT REAGENT. ALTHOUGH THE PT CONDITIONS ARE UNK, THIS EVENT IS BEING REPORTED AS A SERIOUS INJURY BASED ON THE INFO PROVIDED BY THE CUSTOMER THAT TWO DOSE CHANGES WERE MADE. PT CONDITIONS AS A RESULT OF THE DOSE CHANGE ARE UNK.