inforMED
MalfunctionGWJ

ALGO 5

Received Mar 16, 2026 · Event occurred Mar 11, 2026

Report 9612197-2026-00007 · MDR key 24603500

Device

Generic name

Algo 5

Model number

000619

Product problems

  • Electrical /Electronic Property Problem
  • Sparking

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

INITIAL REPORT REF NATUS COMPLAINT #: (B)(4). THE DEVICE HAS BEEN REQUESTED FOR THE RETURN AND EVALUATION. RISK RATING: (B)(4) RISK ANALYSIS, HAZARD 2.0. CAUSE - HIGH LEAKAGE CURRENT. HIGH LEAKAGE CURRENT ON CART CHASSIS. EFFECT (HARM) -ELECTRIC SHOCK - THIRD DEGREE BURN, CARDIAC ARREST, IRREGULAR HEART RHYTHM OR POTENTIAL FATALITIES. RESIDUAL RISK - MODERATE. RISK LEVEL- MEDIUM (11).

Description of Event or Problem

THE MANUFACTURER RECEIVED THE INFORMATION REGARDING THE ALGO 5 DEVICE. THE USER HAS REPORTED FOR "SPARK". IT WAS REPORTED THAT THE AC POWER CORD SUPPLYING THE ALGO 5 DSP BOX ASSEMBLY IS EXPOSED AND SPARKING WHEN THE DEVICE IS POWERED ON. ADDITIONALLY, NO PATIENT OR USER HARM OCCURRED AT THIS TIME.

Description of Event or Problem

THE MANUFACTURER RECEIVED THE INFORMATION REGARDING THE ALGO 5 DEVICE. THE USER HAS REPORTED FOR "SPARK". IT WAS REPORTED THAT THE AC POWER CORD SUPPLYING THE ALGO 5 DSP BOX ASSEMBLY IS EXPOSED AND SPARKING WHEN THE DEVICE IS POWERED ON. ADDITIONALLY, NO PATIENT OR USER HARM OCCURRED AT THIS TIME.

Additional Manufacturer Narrative

AN ADVERSE EVENT WAS INITIALLY REPORTED FOR THIS CASE. UPON DETAILED INVESTIGATION, IT WAS DETERMINED THAT THE COMPLAINT WAS A DUPLICATE/ENTERED IN ERROR. NO NEW ADDITIONAL ADVERSE EVENT WAS OCCURRED AND NO SIGNIFICANT HEALTH IMPACT WAS ASSOCIATED WITH THIS REPORT. THE CASE WAS OPENED IN ERROR.

Additional Manufacturer Narrative

FURTHER INFORMATION AND REVIEW CONFIRMS THAT THE CASE WAS CLOSED IN ERROR. THE SUSPECT DEVICE WAS RETURNED AND SUBSEQUENTLY REPLACED FROM THE CUSTOMER'S SPARE STOCK. NO NEW ADDITIONAL ADVERSE EVENT OCCURRED AND NO SIGNIFICANT HEALTH IMPACT WAS ASSOCIATED WITH THIS REPORT. RISK RATINGS: (B)(4) REV E ALGO 5 RISK ANALYSIS, HAZARD 2.0. CAUSE - HIGH LEAKAGE CURRENT. HIGH LEAKAGE CURRENT ON CART CHASSIS. EFFECT (HARM) -ELECTRIC SHOCK - THIRD DEGREE BURN, CARDIAC ARREST, IRREGULAR HEART RHYTHM OR POTENTIAL FATALITIES. RESIDUAL RISK - MODERATE RISK LEVEL- MEDIUM (11).

Description of Event or Problem

THE MANUFACTURER RECEIVED THE INFORMATION REGARDING THE ALGO 5 DEVICE. THE USER HAS REPORTED FOR "SPARK" .IT WAS REPORTED THAT THE AC POWER CORD SUPPLYING THE ALGO 5 DSP BOX ASSEMBLY IS EXPOSED AND SPARKING WHEN THE DEVICE IS POWERED ON. ADDITIONALLY, NO PATIENT OR USER HARM OCCURRED AT THIS TIME.

Remedial action

  • Replace