inforMED
MalfunctionKTW

2.0MM DRILL BIT WITH DEPTH MARK/QC/140MM

Received Mar 13, 2026 · Event occurred Mar 1, 2026

Report 8030965-2026-02504 · MDR key 24592384

Device

Generic name

Drill Bit

Manufacturer

Synthes Gmbh

Catalog number

323.062

Lot number

59757P4

Product problems

  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: ADDED: H4. H6 COMPONENT CODES: MOST RELEVANT COMPONENT CODE IS G07002 (APPROPRIATE TERM/CODE NOT AVAILABLE) TO CAPTURE NO FINDINGS AVAILABLE DUE TO NO PRODUCT RETURNED. H3, H6: PHOTO INVESTIGATION: THE PRODUCT WAS NOT RETURNED TO DEPUY SYNTHES, HOWEVER PHOTOS WERE PROVIDED FOR REVIEW. THE PHOTO INVESTIGATION REVEALED THAT DRILL BIT Ø2 W/DOUBLE MARKING L140/115 3 HAD BROKEN. THE OBSERVED CONDITION OF THE DEVICE WAS CONSIST WITH A RANDOM COMPONENT FAILURE THAT MAY HAVE BEEN CAUSED BY EXPOSURE TO UNINTENDED FORCE. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE DRILL BIT Ø2 W/DOUBLE MARKING L140/115 3 WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, POTENTIAL CAUSE ATTRIBUTED TO COMPONENT FAILURE AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE. PRODUCT CODE: 323.062. LOT NUMBER: 59757P4. IT WAS ELECTRONICALLY REVIEWED AND NO NONCONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS. THE PRODUCT WAS RELEASED ON: 31 MAR 2025. MANUFACTURING SITE: JABIL BETTLACH. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

IT WAS REPORTED THAT A 2.0MM DRILL BIT WITH DOUBLE MARKINGS BROKE WHILE DRILLING INTO THE BONE.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.