inforMED
MalfunctionLRW

V. MUELLER

Received Mar 12, 2026 · Event occurred Feb 11, 2026

Report 1923569-2026-00002 · MDR key 24580725

Device

Generic name

Scissors

Manufacturer

Steris Corporation

Model number

SA1814

Catalog number

SA1814

Lot number

(10)A24XQS

Product problems

  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

INVESTIGATION OF THE REPORTED EVENT IS IN PROCESS. THE DEVICE IS BEING RETURNED FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem

THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, A PART OF THEIR MAYO DISSECT SCISSORS BROKE OFF. THE PATIENT RECEIVED AN X-RAY AND CONFIRMED NO INSTRUMENT PIECES WERE PRESENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

Additional Manufacturer Narrative

THE DEVICE SUBJECT OF THE REPORTED EVENT WAS RETURNED FOR EVALUATION. DAMAGE (NICKS) ON THE CUTTING EDGE OF THE SCISSOR BLADES WERE OBSERVED; HOWEVER, AN EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED. THE DEVICE HISTORY RECORD FOR THE SUBJECT LOT WAS REVIEWED AND CONFIRMED THE DEVICE WAS MANUFACTURED TO SPECIFICATIONS; NO ABNORMALITIES WERE NOTED. A COMPLAINT REVIEW WAS CONDUCTED FOR THIS LOT OF PRODUCTS AND CONFIRMED THIS TO BE AN ISOLATED EVENT. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.