Received Mar 12, 2026 · Event occurred Dec 8, 2025

Report 3003236702-2026-00004 · MDR key 24579563

Device

Generic name

1.6 K-wire, 60mm, Olive 10mm, 1/pkg

Manufacturer

Medartis Ag

Catalog number

A-5045.41/1

Lot number

67446521

Product problems

  • Break

Patient

60 YR · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE OLIVE WIRE CAN BREAK IN THE SHAFT DURING INSERTION/REMOVING: WHEN THE WIRE SHAFT IS PRE DAMAGED DURING SURGERY. WHEN THE WIRE IS BLOCKED IN THE TIP AREA DURING INSERTION/REMOVING. WHEN FURTHER TORQUE IS APPLIED WHEN THE OLIVE HAS CONTACT WITH AN IMPLANT. BY EXCESSIVE BENDING OF THE SHAFT DURING INSERTION/REMOVING. COMBINATIONS OF THE REASONS ABOVE. VISUAL INSPECTION OF THE THREE PARTS OF THE WIRE SHOWS: THE OLIVE WIRE IS FRACTURED IN TWO PARTS DUE TO MECHANICAL OVERLOAD (FORCE FRACTURE). SEVERAL DAMAGE OF THE SHAFT OF THE WIRE. PLASTIC DEFORMATION OF THE TIP OF THE WIRE. ACCORDING TO MISSING MARKS ON THE OLIVE THE OLIVE HAD PROBABLY NO CONTACT TO AN IMPLANT. THE INVESTIGATION LEADS TO THE ASSUMPTION THAT THE OLIVE WIRE BROKE DUE TO MECHANICAL OVERLOAD (FORCE FRACTURE). HOW AND WHY THIS OVERLOAD OCCURRED CANNOT BE DETERMINED.

Description of Event or Problem

THE TIP OF THE K-WIRE BROKE DURING REMOVAL FROM THE PATIENT. ALL FRAGMENTS WERE RETRIEVED. NO HARM OCCURRED TO THE PATIENT.

Additional Manufacturer Narrative

SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE PROVIDED.

Description of Event or Problem

THE TIP OF THE K-WIRE BROKE DURING REMOVAL FROM THE PATIENT. ALL FRAGMENTS WERE RETRIEVED. NO HARM OCCURRED TO THE PATIENT.